LINEAR GRADUATIONS

Standard Incremental Scales and Masks

  Massstab Blende e Abmessungen

Material:
  • selected floatglass, scribed and broken
  • planarity 10 µm/100 x 100 mm
  • coefficient of expansion 9 µm/°C/m
  • glass thickness for scale 2,0 +0/-0,2 mm
  • glass thickness for mask 1,0 ±0,1 mm´
  • max. length of glass 3300 mm
Graduation:
  • sputtered chrome
  • transmission < 0,3% at 400 - 900 nm
  • reflexion > 60% at 400 - 900 nm
  • gratings 10 : 10 µm, 20 : 20 µm
  • line width ratio 1 : 1
  • max. line tolerance ±10%
  • graduation accuracy < ±3 µm per 1000 mm
  • parallelism graduation to glass edge < 0,3 mm per 1000 mm
  • max. track length 3150 mm
Acceptable defects in the graduation/reference mark area:
  • pinholes < Ø250 µm or several smaller defects whose total area does not exceed this
  • dots < Ø180 µm or several smaller defects whose total area does not exceed this
  • defects visible to the naked eye on the illuminated table are not critical
  • max. 3 defects per 1000 mm
  • minimum distance between two defects 40 mm
Acceptable defects in the surrounding area:
  • dots < Ø0,3 mm
  • max. 6 defects per 1000 mm
          
     
On request:
  • other glass types
  • other glass thicknesses
  • other coatings
  • sliding coat on mask
   
MSD 10 Incremental Scale for HEDS 9200 Module

MSD 10 e Abmessungen

Material:
  • selected floatglass, unpolished
  • planarity 10 µm/100 x 100 mm
  • coefficient of expansion 9 µm/°C/m
  • width of glass 8 mm +0/-0,2 mm
  • glass thickness 0,9 ±0,1 mm
Graduation:
  • line width ratio 1 : 1
  • max. line tolerance ±10%
  • sputtered chrome
  • transmission < 0,3% at 400 - 900 nm
  • reflexion > 60% at 400 - 900 nm
Grating:

100 : 100 µm

Accuracy:

5 or 10 µm/m 

Standard measuring length: (mm)

100, 180, 260, 340, 420, 500, 740, 980, 1160
other lengths on request) 

Scale on profile:

connection scale to aluminium profile with rubber cord

Power supply:

+5 V ±5%, max. 100 mA 

Output:

square wave signals from built-in Subdiving Electronics times 1 

Connection possibilities:

directly to suitable NC or DRO